• Did an incident occur?*
    • Full Definition of 'Incident' (click to expand) 
    • Product Type Citation Definition
      Medical Devices MDR Art. 2(64) Any malfunction or deterioration in the characteristics or performance of a device made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any undesirable side-effect.
      IVDs IVDR Art. 2(67) Any malfunction or deterioration in the characteristics or performance of a device made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any harm as a consequence of a medical decision, action taken or not taken on the basis of information or result(s) provided by the device.

      MDCG 2023-3 Guidance:

      The below information is high-level. Please see MDCG 2023-3 for more specific guidance on these topics. 

      Undesirable Side-Effect

      An ‘undesirable side-effect’ is any unintended medical manifestation as a consequence of using a medical device.

      Malfunction or Deterioration in the Characteristics or Performance of a Device

      If you are unsure what constitutes a ‘malfunction or deterioration in the characteristics or performance of a device’, please refer to pages 11-12 of MDCG 2023-3 for a (non-exhaustive) list of examples. 

      Examples of Incidents

      Examples of incidents can include the following (non-exhaustive list):

      • A device that fails or is losing its ability to achieve its intended purpose when used as instructed in the information supplied by the manufacturer.
      • A deterioration in the characteristics of the device that is related to manufacturing errors, such as sterilisation process failures.
      • A use error due to the ergonomic features of the device, such as a use error caused by a mismatch between the user interface and the physical condition of the intended user.
      • Any inadequacy in the information supplied by the manufacturer, such as insufficient information on how to use, maintain, adjust or calibrate the device, which can lead to a use error.
      • Unclear instructions in the labelling or the manufacturer’s instructions for use, such as information that is not written in a manner that is suitable for/readily understood by the intended user.
      • Incorrect labelling, such as mislabelling of IVD reagents.
      • Undesirable side-effects, such as allergic skin reactions after exposure to nickel or wound therapies.
      • Harm to a person resulting from a medical decision, action taken, or not taken, based on incorrect information or result(s) provided by the device. For instance, incorrect diagnostic results generated by the device that resulted in delayed patient treatment and caused a deterioration in the patient’s health. 
    •  
    • Action Required:

      This situation does not meet the reporting threshold. Therefore, a report to the competent authority is not required. However, this situation, and the assessment of it, must be recorded in your quality management system.

      If new information emerges that changes the nature of the situation, you should re-evaluate whether reporting obligations apply.

    •  
    • Was the incident 'serious', i.e., did the incident directly or indirectly lead, might have led, or might lead to any of the following?
    • Definition and Context on Serious Incidents (click to expand) 
    • Term Citation Definition
      Serious Incident

      MDR Art. 2(65)

      IVDR Art. 2(68)

      Any incident that directly or indirectly led, might have led or might lead to any of the following:

      (a) the death of a patient, user or other person,
      (b) the temporary or permanent serious deterioration of a patient's, user's or other person's state of health,
      (c) a serious public health threat.

      Unanticipated Serious Deterioration in a Person's State of Health

      MDR Art. 87(5)

      IVDR Art. 82(5)

      ...in the event of death or an unanticipated serious deterioration in a person's state of health the report shall be provided immediately after the manufacturer has established or as soon as it suspects a causal relationship between the device and the serious incident but not later than 10 days after the date on which the manufacturer becomes aware of the serious incident.

      MDCG 2023-3 Guidance:

      The below information is high-level. Please see MDCG 2023-3 for more specific guidance on these topics. 

      Unanticipated

      A serious deterioration in a person's state of health, is regarded as ‘unanticipated’ if the condition leading to the deterioration was not considered in the manufacturer’s risk assessment.

      A serious deterioration in a person’s state of health is regarded as anticipated if it was considered in the manufacturer’s risk analysis and documented in the risk management report.

      For a serious deterioration in the state of health, the manufacturer must ensure there is: 

      • documented evidence that a risk management method was used to eliminate or reduce the risk related to these events as far as possible, or
      • the risk is included in the information supplied by the manufacturer to the user, e.g. in the instructions for use.

      Serious Deterioration in the State of Health of a Patient, User or Other Person

      May include:

      1. A life-threatening illness or injury,
      2. Temporary or permanent impairment of a body structure or function,
      3. A condition necessitating hospitalization or prolongation of existing hospitalization,
      4. Medical or surgical intervention to prevent I or II, examples of this can be:
        • Professional medical care or additional unplanned medical treatment,
        • A clinically relevant increase in the duration of a surgical procedure
      5. A chronic disease,
      6. Foetal distress, death, congenital abnormality (including congenital physical or mental impairment) or birth defects..

      Please note that any indirect harm that may occur as a consequence of a medical decision or action taken/not taken on the basis of information or result(s) provided by a device can also lead to serious incidents, including a serious deterioration in a person’s state of health.

      Serious public health threat

      Includes:

      1. The possibility of multiple deaths occurring at short intervals
      2. Events that are of significant and unexpected nature, becoming alarming as a potential public health hazard

      See page 10 of MDCG 2023-3 for a (non-exhaustive) list of examples. 

    •  
    • Action Required:

      As this incident does not meet the threshold for a serious incident, a report to the Competent Authority is not required. However, this incident, and the assessment of its reportability, must be recorded in your quality management system.

      Additionally, multiple incidents of the same or similar nature, including shared root causes or exposure routes, may need to be submitted as a trend report to the relevant Competent Authority in accordance with MDR Article 88 / IVDR Article 83, even where individual incidents were not reported as serious incidents.

    •  
    • Is there a causal relationship between the device and the serious incident, or is a causal relationship reasonably possible?
    • MDCG 2023-3 on Causal Relationships (click to expand) 
    • MDCG 2023-3 advises that the manufacturer must investigate whether there is a causal relationship between the serious incident and their device, or if such a relationship is reasonably possible, i.e., the device cannot reasonably be excluded as a contributory cause of the serious incident.

      In assessing the potential link between their device and the serious incident, the manufacturer should take account of factors such as:

      • clinical or medical plausibility,
      • the opinion of healthcare professionals,
      • the results of the manufacturer’s own preliminary assessment,
      • known information provided in the technical documentation and evidence of previous similar serious incidents,
      • other relevant evidence held by the manufacturer.

      Establishing or identifying the link between the manufacturer’s device and the serious incident may be difficult, especially when there are multiple devices and drugs involved. 

      In complex situations, it should be assumed that the device may have, or potentially could have, contributed to the serious incident, and the manufacturer should therefore be cautious in its evaluation and conclusions.

      In case of doubt, the manufacturer must nevertheless submit the report referred to in Article 87(1) MDR/Article 82(1) IVDR.

    •  
    • Action Required:

      As no causal relationship has been established between the device and the serious incident, a report to the Competent Authority is not required at this time.

      However, the rationale for excluding the device as a contributory cause must be documented in the quality management system, including the factors considered in reaching this conclusion.

      Per MDCG 2023-3, establishing or identifying the link between the manufacturer's device and the serious incident may be difficult, especially when multiple devices and drugs are involved. In complex situations, it should be assumed that the device may have, or potentially could have, contributed to the serious incident, and the manufacturer should therefore be cautious in its evaluation and conclusions. In case of doubt, the manufacturer must report in accordance with MDR Article 87(1) / IVDR Article 82(1).

    • Action Required:

      The serious incident, even in case of doubt of a causal relationship between the device and the serious incident, must be reported no later than 2 days after the awareness date. 

      Per MDCG 2023-3:

      • The ‘manufacturer awareness date’ of the incident is the date when the first employee or representative of the manufacturer’s organisation (i.e. any natural or legal person acting on behalf of the manufacturer) receives information regarding the potentially serious incident.
      • The timelines for reporting serious incidents must be considered as calendar days, meaning the reporting periods include weekdays, public holidays, Saturdays and Sundays.
      • As a general rule, the reporting period begins on the day after the awareness date of a potentially serious incident at 00:00:01 AM.
      • The awareness date (day=0) refers to the date when the manufacturer is first made aware or receives information of the occurrence of the (potentially) serious incident, and not after it has conducted its investigation.

      Please review MDCG 2023-3 for more detailed information, including example reporting timeframes based on manufacturer awareness date.

    • Action Required:

      The serious incident, even in case of doubt of a causal relationship between the device and the serious incident, must be reported no later than 10 days after the awareness date

      Per MDCG 2023-3:

      • The ‘manufacturer awareness date’ of the incident is the date when the first employee or representative of the manufacturer’s organisation (i.e. any natural or legal person acting on behalf of the manufacturer) receives information regarding the potentially serious incident.
      • The timelines for reporting serious incidents must be considered as calendar days, meaning the reporting periods include weekdays, public holidays, Saturdays and Sundays.
      • As a general rule, the reporting period begins on the day after the awareness date of a potentially serious incident at 00:00:01 AM.
      • The awareness date (day=0) refers to the date when the manufacturer is first made aware or receives information of the occurrence of the (potentially) serious incident, and not after it has conducted its investigation.

      Please review MDCG 2023-3 for more detailed information, including example reporting timeframes based on manufacturer awareness date.

    • Action Required:

      The serious incident, even in case of doubt of a causal relationship between the device and the serious incident, must be reported no later than 15 days after the awareness date

      Per MDCG 2023-3:

      • The ‘manufacturer awareness date’ of the incident is the date when the first employee or representative of the manufacturer’s organisation (i.e. any natural or legal person acting on behalf of the manufacturer) receives information regarding the potentially serious incident.
      • The timelines for reporting serious incidents must be considered as calendar days, meaning the reporting periods include weekdays, public holidays, Saturdays and Sundays.
      • As a general rule, the reporting period begins on the day after the awareness date of a potentially serious incident at 00:00:01 AM.
      • The awareness date (day=0) refers to the date when the manufacturer is first made aware or receives information of the occurrence of the (potentially) serious incident, and not after it has conducted its investigation.

      Please review MDCG 2023-3 for more detailed information, including example reporting timeframes based on manufacturer awareness date.

    • Email Results

    • Optional: Please input the below information if you would like it included in the PDF report.

    • What was the awareness date of the incident?
  • Should be Empty: